LOGTAG – TIC20 – TEMPERATURE INDICATOR WITH 20 DAY DISPLAY

TIC 20

The LogTag® TIC20 is a versatile, wide range, multi-trip Temperature Indicator, featuring high resolution temperature readings over a measurement range of -30°C to +60°C (-22°F to +140°F). Enclosed in a robust and durable polycarbonate case, the TIC20 features a real time clock, which provides date/time stamps for each temperature reading.

Using the LogTag® Interface and the freely available companion software LogTag Analyzer, the LogTag® is easily set-up for recording conditions including delayed start, sampling interval, number of readings, continuous or fixed number of readings and configuration of conditions to activate the ALERT indicator. Readings are downloaded using LogTag® Analyzer, which provides facilities for charting, zooming, listing data statistics and allows exporting the data to other applications such as Excel.

The LogTag® TIC20 complies with the relevant international standards for temperature monitoring devices, such as FCC, CE, C-TICK, and RoHS. This not only demonstrates the quality of the LogTag® TIC20, but underlines its suitability for temperature monitoring applications where accuracy and consistency is required

  • At-a-glance summary with ‘OK’/’Reject’ alarm status.
  • 4 freely configurable upper/lower temperature alarms.
  • Review statistical data on device for up to 20 days prior.
  • Download of statistical data via optional interface

Battery life
Minimum storage life of 1 year 6 Months before ‘start’ or minimum storage life of 5 years before ‘start’ (TIC20-W1 & TIC20-W2 only). Monitoring period: 20 days. Minimum accessibility (display) period of 6 months after ‘stopped’.
Case Material
Polycarbonate.
Environmental
IP64
Logging start options
Push button start or specific date & time.
Models
TIC 20
Operating Temperature
-30°C to +60°C (-22°F to +140°F).
Power source
CR2032 3V LiMnO2 Battery (Fixed).
Rated Temperature reading accuracy
Better than ±0.5°C for -10°C to +40°C (+14°F to +104°F). Better than ±1.0°C for -30°C to -10°C (-22°F to +14°F). Better than ±1.0°C for +40°C to +60°C (+104°F to +140°F). Actual performance is typically much better than the rated values. Accuracy figures can be improved by recalibration.
Recording cycle
Minimum & maximum temperature for each of the 20 days. 1st activation of each alarm for each day including trigger time and duration of each excursion. Please note the TIC20 does not store each individual temperature reading taken.
Recording indication
Alarms are displayed on a custom LCD.
Sampling Interval
1-15 minutes.
Size
73mm(H) x 54.5mm(W) x 8.6mm(T).
Start-up delay
Start delay of 1-99 minutes.
Storage Method
Temperature Range: -30°C to +65°C (-22°F to +149°F).
Weight
30g

Comparison (0/2):

Click on compare button to add products to comparison.